| Determine Kc |
Honors Chemistry |
background [prelab assignment (15 points) due at the begining of lab]
what is Beer's Law ? calibration curve (relate to Beer's law; hint: vitamin C lab) ?
fill-in the below ICE table in regards to the below reaction
[Fe3+] [SCN-] [FeSCN2+] [inital]5.0 mM 10.0 mM 0 [change] [equilibrium]2.0 mM and calculate the value of Kc based on the information in the above ICE table.
msds on nitric acid, HNO3
view screen cast on the use of the colorimeter prior to the lab
purpose
materials (per 2 groups @ a table)
colorimeter; cuvette; 100 mL volumeric flask
100 mL, 200.0 mM iron (III) nitrate; 50 mL, 2.0 mM potassium thiocyanate
2 mM iron (III) nitrate (dilute 200 mM solution)
150 mL, 0.1 M HNO3 - use as the solvent
methods
student design; delegate your work - do calculations on how to prepare the below solutions for tables 1 - 3 prior to the lab
table 1. concentration of __ in 100 mL mixture for calibration curve. Use 200 mM Fe(NO3 )3 , 2 mM KSCN, and 0.1 M HNO3 to prepare these solutions. A group makes a solution, label it, and share with your classmates @ all tables.
std [Fe3+ ] (mM) [SCN- ] (µM) [FeSCN2+] absorbance 1 40 2002 40 1503 40 1004 40 50zero 40 0table 2. concentration of __ in 10 mL mixture - to determine Kc . Use 2 mM Fe(NO3 )3 (dilute 200 mM solution to make 2 mM solution), 2 mM KSCN, and 0.1 M HNO3 to prepare these solutions and share with your classmates @ the same table.
[Fe3+ ] (mM) [SCN- ] (mM) absorbance 1.0 0.1 1.0 0.5 1.0 1.0table 3. concentration of __ in 10 mL mixture - to determine the effect of [Fe3+ ] on [FeSCN2+ ]. You'll have to determine how to prepare these solutions, and share with your classmates @ the same table.
[Fe3+ ] (mM) [SCN- ] (mM) absorbance 0.1 0.1 0.2 0.1 0.5 0.1 1.0 0.1 2.0 0.1
use of colorimeter (vodcast; assume only FeSCN2+ absorbs light)
- turn-on Xplorer data collection device
- insert colorimeter into Xplorer device
- open lid, then insert small glass vial (cuvette) filled with "zero" (see table 1, above) into colorimeter
- close lid; should display %T ~ 100 or Abs ~ 0 for all wavelengths; press "+" on Xplorer - to cycle thru various setting (red, green, blue, & orange, where reading % T or Abs
- if %T not ~ 100% or ABS not ~ 0 , then press green button on colorimeter / wait till lights goes off else goto next step
- replace cuvette with samples; press "+" on Xplorer - to cycle thru various setting; record data (i.e. absorbance at different wavelengths); repeat with other samples
- DO NOT PRESS ANY OTHER BUTTON DURING YOUR MEASUREMENT
data analysis
calibration curve; ICE table
Content of lab report (30 points; no lab report revision)
data analysis
calibration curve - based on data in table 1
determine Kcbased on the ICE table & data in table 2; show all calculations
summary tables / graph:
- Kc as a function of the initial [Fe3+] & [SCN- ]; (based on table 2)
- make / label a graph of experinental data (curve 1): initial [Fe3+] versus equilbrium [FeSCN2+] (based on table 3)
statistics / conclusions:
- is the value of Kc truly a constant (i.e. is the value of the Kc the same under all (different) initial conditions) ? use / refer to the appropriate statistics to support your conclusion.
- describe / rationalize the relationship between initial [Fe3+] versus equilbrium [FeSCN2+]
resources:
site 1; site 2